2 Hardware
2
Instructions for Use
Resting & Stress ECG with seca ct330/331 and seca diagnostic 5.9
2.8 Attaching the recorder to the patient
2.8.1 Positions of the electrodes
Thoracic wall resting and stress ECG, standard according to Wilson
V1 (C1)
red
4th intercostal space at the right sternal border
V2 (C2)
yellow
4th intercostal space at the left sternal border
V3 (C3)
green
On the left on the 5th rib between C2 and C4
V4 (C4)
brown
5th intercostal space on the left midclavicular line
V5 (C5)
black
On the left on the anterior axillary line, on the level of C4
V6 (C6)
purple
On the left on the midaxillary, on the level of C4
Resting ECG extremities
R
red
Right arm
L
yellow
Left arm
F
green
Left leg
N
black
Right leg
Stress ECG (lying or sitting position): extremities
R
red
On the right below the collarbone
L
yellow
On the left below the collarbone
F
green
On the left above the hip
N
black
On the right above the hip
Stress ECG (sitting position): extremities
R
red
Attach to the deltoid muscle on the right
L
yellow
Attach to the deltoid muscle on the left
F
green
9th rib left
N
black
9th rib right
Fig.
5: C1 to C6
Fig.
6: R, L, F, N