A01350693 Rev:P ECO: 7476 Date: 04/14/08 42
Recommended Weekly Testing
Testing Wandering Patient Monitoring with Lock
Patient Escort Feature Test
Enter the monitoring zone with a transmitter on your ankle. The red LED will turn on and the door will quietly lock.
Now enter the primary reset code into the keypad. The red LED will turn green, the door will unlock and you will be
able to open the door without the alarm sounding.
Anti-Tailgate Feature Test
Now close the door. The green LED will turn red and the door will immediately relock.
Delayed Egress Feature Test
With the door locked, apply pressure on the opening hardware of the door. The 135DE Exit Panel will begin to alarm
and the red LED will remain on. The LED will remain red and the audible alarm sound will be short beeps. After 15
seconds, the audible alarm will become a continuous tone, the red LED will be on, and the door will release. Open the
door. The LED will begin flashing red/green/red/green and the audible alarm will change to the seagull sound. Now
close the door and enter the primary reset code. The panel will stop alarming, the door will relock and the red LED will
be on. (The transmitter is still in the monitoring zone).
Remote Keypad Test
If a Remote Keypad has been installed with the 135DE Exit Panel, repeat the steps above using the Remote Keypad.
Push Button Test
If a remote Push Button has been installed with the 135DE Exit Panel, pass through the door from the Exit Panel side to the
Push Button side using the reset code on the Exit Panel keypad. Close the door. Depress the Push Button. Open the door. No
alarm condition will be activated
and the LED will turn green.
Advanced Security Mode Test
Bring a transmitter into the monitoring zone. The green LED will turn red; the yellow LED will blink. Now enter the
primary reset code. The red LED will flash green momentarily indicating a valid code was entered while a monitored
resident was nearby. Next, try to gain access via the Push Button if one is installed on your system. With a transmitter
in range, the door will remain locked and an audible alarm will sound. Only by entering the Secondary Reset Code
(*567 or equivalent) will access be granted in this situation.
Recommended Monthly Testing
Fire Alarm Release Feature Test
Place the fire alarm system into the alarm mode and verify that all locks release. The 135DE Exit Panels should be in the fire
alarm release alarm condition.
Recommended Annual Service
Battery Replacement
The battery should be replaced annually with a rechargeable 9VDC NI-MH battery (Secure Care Part # B15360501).
SECTION 11 TESTING