3.2
3.2 Setting time and date
The time, date and date format (Day/Month/Year or Month/Day/Year) may be set by
following the procedure in Table 3.2. History log entries will have an incorrect time and
date stamp prior to setting the clock time and date. The time and date must be set
correctly before setting up auto calibration. The time and date are retained for at least
two days while the analyser is switched off, the configuration and all calibrations are
retained indefinitely.
Table 3.2 Setting time and date
L MENU to obtain top level menu
CALIBRATE/
SETUP
ALARMS/FAULTS
L < ENTER
SET ALARM/ASSIGN
DISPLAY/UTILITY
L ? < ENTER
ENTER PASSWORD
0000
To change the value of a digit
L > or ?
To change to another digit L = or <
When the number shown is correct L ENTER
UTILITY 1
UTILITY 2
L ENTER
CLOCK/NEW PASS
COMMS/WINDOW
L ENTER
TIME = 12:54
DATE = 10/12/95
Example of temporary display.
CHANGE DATE,TIME
YES/NO
L ENTER
SET CLOCK
SET YEAR 0000
To change the value of a digit
L > or ?
To change to another digit L = or <
When the value shown is correct L ENTER
SET MONTH
00"
SET DAY 00"
SET HOUR 00"
SET MINUTE 00"
SELECT FORMAT
DDMMYY/MMDDYY
To select Day/Month/Year format L ENTER
To select Month/Day/Year format L <
ENTER
TIME = 12:57
DATE = 10/12/95
Example of temporary display.
CLOCK/NEW PASS
COMMS/WINDOW
To return to measurement display L
MEASURE