5.4
Table 5.2 Allocation of relays
L MENU to obtain top level menu
CALIBRATE/SETUP
ALARMS/FAULTS
L < ENTER
SET ALARM/ASSIGN
DISPLAY/UTILITY
L < ENTER
ENTER PASSWORD
0000
To change the value of a digit L > or ?
To change to another digit L = or <
When the number shown is correct L ENTER
RELAYS
ANALOGUE
L ENTER
SELECT RELAY
1.3 ASSIGNED 8
or
1.3 UNASSIGNED
8
The ASSIGNED and UNASSIGNED informs the
user if the relay has been previously assigned.
L > or ? to select the required relay
then
L ENTER
RELAY ASSIGNMENT
1.3 EDIT/CLEAR
or
1.3 ASSIGN?
Y/N
EDIT or CLEAR an existing relay assignment or
ASSIGN a new one. To change to selection
L
=
or <. When correct selection is highlighted
then
L ENTER.
1.3I1 Oxygen AL1
HI 10.000 % Y/N
or
1.3 CAL IN PROG
YES/NO
or
1.3 FAULT
YES/
NO
The display will now show a series of items
which may be selected to operate the relay
being configured.
L = ENTER to allocate the item to the relay.
L ENTER to move to the next item.
If the item being offered for selection is an alarm
the alarm setpoint will be displayed, an H or L
will also be displayed indicating wether it is
configured to be a High or Low alarm.
SELECT RELAY
1.3 ASSIGNED
8
or
1.3 UNASSIGNED8
L
> or ? to select another relay then L
ENTER
or
L MEASURE to return to measurement
display

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO 4900 and is the answer not in the manual?

Servomex SERVOPRO 4900 Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO 4900
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals