7.2
7.4 Toxic Samples - routine leak test
WARNING
If toxic samples are being analysed it is essential to check the analyser and
associated sample lines/system for leaks (every 6 months is recommended
period). MAX pressure that may be applied to each module is 8 psig, however,
this must be applied slowly to both the inlet and outlet simultaneously to avoid
damage to the measuring sensors.