EasyManua.ls Logo

Si-tex Explorer Nav Pro - Transducer Setup

Si-tex Explorer Nav Pro
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
66
Chart Plotter - User Manual
In Bottom Lock Mode the Fish Finder automatically tracks the range around the bottom
specied by the Bottom Range value.
If the range is set to Manual, it can be changed by zooming in/out on the Fish Finder page.
Zooming in/out can be done with ZOOM IN and ZOOM OUT keys of the chart plotter, and with
the wheel of the remote control.
In the sensitivity page you can also set the sensitivity time constant and the surface noise lter
Sensitivity Time Constant: it is a time varying gain curve, which attenuates the sonar receiver
gain in shallow water, increasing the gain gradually as the depth increases. This is for the
purpose of ltering out surface clutter.
Surface noise lter: implements an advanced time varying lter that allows suppressing the
surface clutter e󰀩ciently while leaving sh targets visible.
18.6 Transducer setup
Fig 18.6 – Transducer setup page
On the Transducer setup page you can customize:
Keel o󰀨set: the transducer depth o󰀨set from the surface. This makes it possible to
measure depth from the surface instead of from the transducer's location.
Calibrate sound speed: allows calibrating the value of sound speed in the water,
depending on the water temperature and salinity.
Calibrate water speed: calibrates the value of water speed sensor. The calibration
value, in the range between -10% to +10%, will be applied to the water speed from

Table of Contents

Related product manuals