31
8011952/YWL2/3-0/2016-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
PRODUCT DESCRIPTION 2
2.3 Device configuration
The device components required for a measuring system depend on the respective appli-
cation conditions. The following Table should serve to assist you in your selection.
Sender/receiver unit, reflector/scattered light receiver, flange with tube
Voltage and purge air supply
NOTE:
The reflector/scattered light receiver may not project from the flange tube. The
maximum possible wall and insulation thickness is therefore limited to the
values stated in the Table.
Distance
flange -
flange
Maximum
wall and
insulation
thickness
Sender/
receiver
unit
Reflector/
scattered
light
receiver
Flange with tube
Line for reflec-
tor/scattered
light receiver
Sender/receiver
unit
Reflector/scattered light
receiver
0.5 ... 3
m
40 mm
DHC-T
DHC-R0
Flange with tube
k100
NL 130/240/500
mm
Flange with tube k100
NL 110 mm
x
2.5 ... 8
m
270 mm DHC-R1
Flange with tube k225
NL 350 mm
Internal duct
pressure
Distance MCU - sender/receiver
unit or reflector/scattered light
receiver
Connection and supply components
Purge air Voltage
up to +2 hPa max. 3 m
MCU-P + purge air hose DN 25
(on sender/receiver unit and reflector/scattered light receiver DHC-R0)
or purge air hose DN 40 (on reflector/scattered light receiver DHC-R1)
> +2 hPa
Optional external purge air unit + adapter 40-25
(on sender/receiver unit)
MCU-N