59
8011952/YWL2/3-0/2016-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
Fig. 34: Assembly on a dust-free path with zero tube
Using the associated connection lines, connect the sender/receiver unit to the MCU and
the reflector/scattered light unit to the sender/receiver unit.
Connect the MCU to the supply voltage.
Start the SOPAS ET program and connect to the measuring system (see “Connection to
the device via USB line”, page 55).
Enter the Level 1 password (see “Password and operating levels”, page 84).
Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”.
Fig. 35: SOPAS ET menu: DH C200/Maintenance/Maintenance
Clean the optical surfaces on sender/receiver unit and reflector (see “Maintenance on
the sender/receiver unit”, page 94 and see “Maintenance on the reflector”, page 98).
Wait approx. 30 min before starting the following work (measuring system must have
reached operating temperature).
A
Ø76
X
X: Dimensions between flange and tube according to the
type of reflector/scattered light receiver (see “Reflector/
scattered light receiver”, page 115, see “Flange with tube”,
page 117)
Sender/receiver unit
Reflector/scattered light receiver
Adjusting stand