START-UP AND PARAMETER SETTINGS 4
Fig. 34: Assembly on a dust-free path with zero tube
▸ Using the associated connection lines, connect the sender/receiver unit to the MCU and
the reflector/scattered light unit to the sender/receiver unit.
▸ Connect the MCU to the supply voltage.
▸ Start the SOPAS ET program and connect to the measuring system (see “Connection to
the device via USB line”, page 55).
▸ Enter the Level 1 password (see “Password and operating levels”, page 84).
▸ Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”.
Fig. 35: SOPAS ET menu: DH C200/Maintenance/Maintenance
▸ Clean the optical surfaces on sender/receiver unit and reflector (see “Maintenance on
the sender/receiver unit”, page 94 and see “Maintenance on the reflector”, page 98).
▸ Wait approx. 30 min before starting the following work (measuring system must have
reached operating temperature).
A
Ø76
X
X: Dimensions between flange and tube according to the
type of reflector/scattered light receiver (see “Reflector/
scattered light receiver”, page 115, see “Flange with tube”,
page 117)
Sender/receiver unit
Reflector/scattered light receiver
Adjusting stand