Configuring with the WBM
7.3 The menu items of the WBM
SIMATIC RF600
84 Configuration Manual, 11/2018, C79000-G8976-C386-07
Procedure for determining the activation power
Note
Do not optimize the activation power during operation
We recommend that you do not optimize (
) th
e activation power during operation since this
can disturb operation.
Select the required read point and antenna, enter the required measuring range values
and start the measurement.
Select a transponder from the list.
power to the "Radiated power" input box of the "Settings - Read
In this area, you can select the read point and antenna whose optimum activation power you
want to measure.
With the "Start/Stop measurement" (
/ ) and "Delete display" ( ) buttons, you can control
the power measurement.
By clicking the
button, a new recording is started. Cyclic reads are performed on the
selected antenna and the measured values obtained are displayed. Any existing measured
values from a previous recording are deleted. By clicking the
button, the recording is
stopped. Clicking the
button deletes the currently displayed measured values.
Using the input boxes in this area, you can influence the measurement. The parameters
contained in this area are intended for trained users. For untrained users, we recommend
using the default settings.
● Power from ... to ...
Specifies the range of values (dBm value) within which the measurement is made. The
measurement starts at the "from" value and ends automatically as soon as the "to" value
is reached.
● Increment
Specifies the dB value by which the radiated power is increased step-by-step during the
measurement.