The adapters need to be fastened with 35 Nm in case they get loosed during installation.
8.2.4.2 Electrical connections with conduits
Electrical connections
Door screws
Shielding plate
Procedure
1. Undo the six screws and open the door of the wall-mounted device.
2. Remove the shielding plate on the right-side panel. To do this loosen the screws.
3. Strip the end of signal cable to a length of approx. 1 cm.
4. Remove the blind plug.
5. Insert the signal and power cable
6. Properly install a conduit connection. Thread Dimension NPT1. In case purging is used, make
sure that the conduit is gas tight sealed.
To see the terminal assignment go to chapter AUTOHOTSPOT
Connecting the device
8.2Electrical connections
Wall-mounted device
110 Operating Instructions, 07/2023, A5E31930403-AB