EasyManua.ls Logo

Siemens SIPROCESS GA700 - [2.09] Settings > Inputs;outputs; [2.09.1] Analog inputs

Siemens SIPROCESS GA700
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test setup for leakage test with ULTRAMAT 7 high-temperature
To check for leaks, we recommend the following test setup:
X
Z
Y
1
2
4
3
Shut-o󹪜 valve
Pressure regulator
Manometer (gauge pressure)
1 Sample gas inlet
2 Sample gas outlet
3 Unscrewing blanking plugs
4 Unscrewing blanking plugs
Figure9-6 Recommended test setup for leakage test with ULTRAMAT 7 high-temperature
9.6.5 Checking OXYMAT 7 for leaks
Procedure: Devices with pipes
1. Close the reference gas inlet.
2. Connect a pressure measuring instrument (manometer) to the sample gas outlet.
3. Carefully apply pressure of approximately 3000 hPa (absolute pressure) to the measuring
inlet.
4. Then shut o󹪜 the measuring inlet.
5. Wait for approximately 5 minutes until the test gas that 󹪞owed in has reached the ambient
temperature and the pressure remains stable.
Note the pressure.
6. Wait an additional 5 minutes and note the pressure again.
If the pressure change is less than 5 hPa, the sample gas path is adequately tight.
This value is based on the tare volume of the measuring module 7 ml and the volume of the
pressure measuring device of 23 ml.
Commissioning
9.6Checking gas paths for leaks
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 137

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Related product manuals