Note
Device variant Cleaned for O
2
In the device version with the order sux -B06 (Cleaned for O
2
), all parts wetted by the sample
gas are cleaned and absolutely grease-free.
NOTICE
Formation of condensation in the device
Condensation can be caused by a high dierence in temperature between transportation and
storage and the installation location.
• Before taking the device into operation let the device adapt for several hours in the new
environment.
9.2 General information on commissioning
Note
Ambient temperature
Ensure that the ambient temperature is within the permissible range during operation. The
corresponding information is available in the section "Technical specications".
The technical specications of the device can deviate outside these temperature limits.
Operability of the device is no longer ensured at deviations of more than 5K.
Note
Vibrations at the installation site disturb the measured signal
When you use OXYMAT7 and/or ULTRAMAT7 analyzer modules, vibrations, especially in the
frequency range from 6Hz to 15Hz, can inuence the measured signal.
• Select an installation location which is vibration-free.
• If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not correspond to the fundamental of the
installation site nor one of its harmonic multiples.
– You must also ensure that the measuring frequency diers from half the basic frequency
of the vibration.
Commissioning
9.2General information on commissioning
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 123