If you are using SIPROCESS GA700 devices in critical processes, observe the following
information:
Monitor the limits externally via the control system.
If an option module 2.1 is installed, read out the device status via the digital outputs of the
option module.
Use an available manual measuring range switchover instead of the autoranging.
Only use the "Total" parameter with the "O󹪜" setting for the measuring range switchover.
If you use the "Correction of cross-interference" and "External pressure compensation"
functions, store suitable substitute values.
Only access the measured value via an analog output.
3.4 Measuring high oxygen concentrations
Note
Measuring high oxygen concentrations
For the measurement of high O
2
concentrations in excess of 25%, it is absolutely necessary that
all parts that come in contact with the sample gas are free of oil and grease and do not contain
any particles that could react with O
2
.
We highly recommend you use specially cleaned analyzer modules with the "Cleaned for O2"
option for such measurements.
General information
3.4Measuring high oxygen concentrations
Wall-mounted device
20 Operating Instructions, 07/2023, A5E31930403-AB

Table of Contents