Inuence variables
Accompanying gases Zero point deviation (cross-sensitivity) in accordance with Table A.1 of EN
61207-3
Supply voltage Analyzer module 24VDC
(-5%/+10%)
< 0.1% of the current span
Basic device
85VACto 264VAC / 47Hz to
63Hz
< 0.1% of the current span
Table 13-14 OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport -30…+70°C (-22 ... 158 °F)
Max. permissible ambient temperature surround‐
ing the basic device during operation
1)
0…50°C (0 ... 122 °F)
Ambient humidity (relative humidity) during stor‐
age, transport or operation
2)
<90%
1)
Restriction for installing an ULTRAMAT7 analyzer module (combined operation): 5…45°C (41 ... 113 °F)
2)
Avoid condensation on the installed components!
Table 13-15 OXYMAT 7: Gas connections
Gas connections
With hoses Connecting socket for 6 mm hose
With pipes Connecting socket for 6 mm pipe
Technical specications
13.3Technical specications of analyzer modules