Table 13-35 CALOMAT 7: Climatic conditions
Climatic conditions
Storage and transport -30…+70°C (-22 ... 158 °F)
Max. permissible ambient temperature surround‐
ing the basic device during operation
1)
0…50°C (32 ... 122 °F)
Ambient humidity (relative humidity) during stor‐
age, transport or operation
2)
<90%
1)
Restriction for installing an ULTRAMAT7 analyzer module (combined operation): 5…45°C (41 ... 113 °F)
2)
Avoid condensation on the installed components
Table 13-36 CALOMAT 7: Gas connections
Gas connections
Connection ttings Pipe diameter 6mm (outer diameter)
Table 13-37 CALOMAT 7: Parts in contact with sample gases
Materials of wetted parts
Gas connection Stainless steel (Material No. 1.4571)
Clamping rings and union nut (set) Stainless steel (Material No. 1.4401)
Sample gas pipes Stainless steel (Material No. 1.4404)
Sensor mounting block Stainless steel (Material No. 1.4571)
Sensor Si, SiO
x
N
y
, Au, epoxy resin, glass
Gasket, contained in the sensor module Peruorelastomere FFPM
Technical specications
13.3Technical specications of analyzer modules
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 173