EasyManua.ls Logo

Siemens SIPROCESS GA700 - Page 69

Siemens SIPROCESS GA700
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Remove the equipotential bonding cable from the blank plate of the respective slot.
Note: The cable for equipotential bonding is no longer required after the analyzer
module is installed.
Figure7-3 Unlatching plug-in connectors at the cable equipotential bonding
NOTICE
Damage to the device through pulling the latched cable
Cables for equipotential bonding are equipped at their two ends with latching plug-in
connectors. Pulling on a latched cable can cause the plug-in connector to be torn o󹪜 or its
contacts to be damaged.
Before pulling out the plug connector, loosen the latch by pressing with your index 󹪝nger.
4. Loosen the eight Torx screws and remove the blank plate at the corresponding slot.
5. Check the 󹪞at seal for damage.
Replace the 󹪞at seal if it is damaged or porous.
6. Swivel the analyzer module forward and insert it into the wall housing .
7. Secure the analyzer module in the slot as follows:
Use eight Torx screws on the base of the housing with a torque of 0.8Nm.
Use two Torx screws on the rear panel with a torque of 2.5Nm.
Installing / removing and connecting analyzer and option modules
7.4Installing analyzer modules, standard version
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 69

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Related product manuals