4. Loosen the eight Torx screws ③ and remove the two blank plates at slots 1 and 2.
5. Check the at seal ④ for damage.
Replace the at seal if it is damaged or porous.
6. Swivel the analyzer module ⑥ forward and insert it into the wall housing ①.
7. Secure the analyzer module ⑥ in the slot as follows:
– Use eight Torx screws on the base of the housing ③ with a torque of 0.8Nm.
– Use two Torx screws on the rear panel with a torque of 2.5Nm.
8. Connect the gas lines, see section "Gas connections (Page91)".
Installing / removing and connecting analyzer and option modules
7.5Installing OXYMAT 7, high-temperature version
Wall-mounted device