3. Remove the following cables:
– Power cable to the analyzer module printed circuit board and power supply unit
– 12-pin ribbon cable (CAN) to the analyzer module printed circuit board and the
processing unit
– If applicable, 26-pin ribbon cable to the analyzer module printed circuit board or the
analog output adapter card and the option module OM 2.1 or OM 2.2
4. Loosen the eight Torx screws on the base of the housing ⑤ and the two Torx screws on the
rear wall.
5. Swivel the analyzer module ⑥ forward and remove it from the wall housing.
6. If you are not installing and connecting a new analyzer module:
Fasten the blank plate over the respective slot using the eight Torx screws ⑤ with a torque
of 0.8 Nm.
7. Close the door of the housing with the six screws ②.
Tighten the screws ② on the door with a torque of 3.5Nm.
Installing / removing and connecting analyzer and option modules
7.8Removing analyzer modules, standard version
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 77