4. Remove the following cables:
Power cable to the analyzer module printed circuit board and power supply unit
12-pin ribbon cable (CAN) to the analyzer module printed circuit board and the
processing unit
If applicable, 26-pin ribbon cable to the analyzer module printed circuit board or the
analog output adapter card and the option module OM 2.1 or OM 2.2
5. Loosen the eight screws on the housing 󹪞oor and the two Torx screws on the rear wall.
6. Swivel the analyzer module forward and remove it from the wall housing.
7. If you are not installing and connecting a new analyzer module:
Fasten the blank plates over the two slots using eight Torx screws in each case with a
torque of 0.8Nm.
8. Close the door of the housing with the six screws .
Tighten the screws on the door with a torque of 3.5Nm.
Installing / removing and connecting analyzer and option modules
7.9Removing OXYMAT 7, high-temperature version
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 79

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals