Performing installation
① Option module OM 2.1 ⑤ Screws for option module (2 units)
② Screw, printed circuit board holder ⑥ Electronics base plate
③ Printed circuit board holder ⑦ Screw, electronics base plate
④ Screw, electronics base plate ⑧ Shielding plate signal lines
Figure7-14 Installing option module, example OM 2.1
1. Loosen the screws ④ and ⑦. Remove the electronics base plate ⑥.
2. Loosen the screw ② and remove the printed circuit board holder ③.
3. Remove the blanking cover plate for option modules.
4. Insert the option module ① into the electronics base plate ⑥ at the respective slot.
Fasten the option module with the two screws ⑤ with a torque of 1.3Nm.
5. Insert the printed circuit board holder ③ again.
Tighten the screw ② with a torque of 2.0Nm.
6. Fit the electronics base plate ⑥ with the two screws ④ and ⑦ at the rear panel and side
panel of the wall-mounted device.
Tighten the screws ④ and ⑦ with a torque of 2.0Nm.
Installing / removing and connecting analyzer and option modules
7.11Installing option modules
Wall-mounted device
84 Operating Instructions, 07/2023, A5E31930403-AB