Diagnostic Action Comments
9 Oil: Density An error in the oil density calculation was detec‐
ted and the measurement accuracy can be af‐
fected. Check if device is configured properly,
investigate DSL error codes to determine pre‐
cise root cause.
Only hydrocarbon
10 Oil: Viscosity An error in the oil viscosity calculation was de‐
tected and the measurement accuracy can be
affected. Check if device is configured properly,
investigate DSL error codes to determine pre‐
cise root cause.
Only hydrocarbon
11 Oil : Spec. gravity An error in the oil gravity calculation was detec‐
ted and the measurement accuracy can be af‐
fected. Check if device is configured properly,
investigate DSL error codes to determine pre‐
cise root cause.
Only hydrocarbon
14 Path 1: No signal Potential reasons: Process conditions, wrong
installation, defect in electronics, cabling or sen‐
sor. It may be necessary to adjust either instal‐
lation or configuration or to replace compo‐
nents. Provide DSL error codes to service per‐
sonnel for further information.
15 Path 2: No signal Potential reasons: Process conditions, wrong
installation, defect in electronics, cabling or sen‐
sor. It may be necessary to adjust either instal‐
lation or configuration or to replace compo‐
nents. Provide DSL error codes to service per‐
sonnel for further information.
19 Electronics temp. The internal DSL temperature is outside the
specified operating temperature range. Verify
that the ambient temperature is within specified
limits. Working outside temperature limits can
damage electronics and reduce product lifetime.
22 Density calculation Measurement of Process Density has failed.
This could be due to errors in attached cabling or
measurement equipment, defect sensor fron‐
tend electronics or configuration errors on chan‐
nels 3, 4, 5 or 6. Investigate DSL error codes to
determine precise root cause and required ac‐
tions. For processes with a constant density, the
density can alternatively be based on a fixed
user specified value.
23 Medium temp. calc. The measurement of Process Temperature has
failed. This could be due to errors in attached
cabling or measurement equipment, defect sen‐
sor frontend electronics or configuration errors
on channels 3, 4, 5 or 6. Investigate DSL error
codes to determine precise root cause and re‐
quired actions. For processes with a constant
temperature, the process temperature can alter‐
natively be based on a fixed user specified val‐
ue.
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD 103