9.8 Fault codes and corrective actions
Alarms and system messages support both Siemens standard alarm classes and NAMUR
status signals.
In the following tables the alarm IDs (identification numbers) are listed along with possible
causes and directions for corrective action.
Sensor diagnostic events
ID/
Icons
Diagnostic Action Comments
5 DSL voltages DSL-internal voltage limits exceeded. Please
check the power connection and repower the
device. If the alarm still exists, call service per‐
sonnel. Electronics may be damaged and it may
be necessary to replace DSL electronics.
6 DSL Storage A malfunction in the DSL-internal memory was
detected. Parameters are either not initialized
correctly or stored correctly. Repower the de‐
vice and check parameter values. If alarm still
exists, call service personnel. Internal data stor‐
age may be damaged and it may be necessary
to replace DSL electronics.
7 Flow measurement An error in the flow measurement system detec‐
ted and the measurement accuracy can be af‐
fected. Try to repower the device. If alarm still
exists, call service personnel. It may be neces‐
sary to replace the DSL electronics or firmware.
8 DSL internal Try to repower the device. If alarm still exists,
call service personnel. It may be necessary to
replace the DSL electronics or firmware.
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
SITRANS FST030 (Modbus)
102 Operating Instructions, 08/2018, A5E37367739-AD