C.1.11 Parameters accessed via drop-down menus
Click on Device, View or Diagnostics to open the associated drop-down menus.
Drop-down menus
Table C-1 Device menus
Device menus Description
Communication Path Shows the communication interface (HART modem)
Communication (Modbus) Shows the communication interface (Modbus interface)
Download to Device Downloads all writable parameters to the device
Upload to PC/PG Uploads all parameters from the device to the parameter
table
Update Diagnostic Status Reads current diagnostic status from the device and up‐
dates the diagnostic status icon
Set Address Sets the HART polling address
Wizard - Quick Start Guide for a quick commissioning
Wizard - Clamp-On Configuration... Guide for a configuration of clamp-on sensors
Wizard - Inline Configuration... Guide for a configuration of inline sensors
Trim Signal Output (online dialog) Calibration of current output (channels 2 to 4)
Totalizer (online dialog) Controlling totalizers 1, 2 and 3
Multipoint calibration The multipoint calibration allows the user to set a calibra‐
tion curve, if deemed necessary, at various flow rates
across the range of operation
FW update A FW update is a firmware update. This function allows the
user to install firmware that has been placed on the SD-
card
Maintenance (online dialog) Setup of maintenance functions
Suppress Alarms Suppresses individual alarms
Audit Trail Lists parameter changes, FW updates, and alarm history
logs
Simulation (online dialog) Simulation of process values, alarms, and inputs/outputs
(channels 2 to 4)
Access Management Possibility to upgrade access level from User to Expert and
to change PIN code for "expert" level. Also gives the option
to change the default access level.
Reset (online dialog) Resets device to default settings or restarts device
SIMATIC PDM
C.1 Commissioning with PDM
SITRANS FST030 (Modbus)
348 Operating Instructions, 08/2018, A5E37367739-AD

Table of Contents