ID/
Icons
Diagnostic Action Comment
238 Pressure Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Lower warning limit"
239 Pressure Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Lower alarm limit"
244 Kinematic viscosity Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Upper alarm limit"
245 Kinematic viscosity Value above warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Upper warning limit"
246 Kinematic viscosity Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Lower warning limit"
247 Kinematic viscosity Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Lower alarm limit"
248 Rate of change Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Upper alarm limit"
Only hydrocarbon
249 Rate of change Value above warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Upper warning limit"
Only hydrocarbon
252 Standard kinematic viscosity Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Upper alarm limit"
Only hydrocarbon
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD 117