Workflow Information
107
Topogram
Select single scan protocol or all scan protocols you
want to modify in this subsection.
In the subsection topogram the following changes can
be made:
–mA
Set the mA value for the topogram.
– Topogram length
All available topo lengths are listed in the dropdown
list.
–Tube position
Set the tube position either to Top or Lateral.
–kV
Can be defined individually to the Scan value.
– Auto transfer modes
Sends images automatically
–Auto Viewing and Filming
Loads images automatically into the Viewing or Film-
ing card.
–API
Select one of the predefined breathing commands
from the dropdown list.
– Scan direction
Select Head to Feed or Feed to Head from the drop-
down list.