428
Specials
For SOMATOM Emotion 6-slice configuration:
A scan range of 12/14 cm will be done in 9.84/13.67
sec.
Emotion 6 Head Neck
kV 130 130
Effective mAs/
Quality ref. mAs
250 150
Rotation time 1.0 sec. 1.0 sec
Acquisition 6 x 3.0 mm 6 x 2.0 mm
Slice collimation 3.0 mm 2.0 mm
Slice width 6.0 mm 5.0 mm
Feed/Rotation 15.3 mm 12.0 mm
Pitch Factor 0.85 1.00
Increment 6.0 mm 5.0 mm
Kernel H31s B50s
CTDI
Vol
55.00 mGy 16.35 mGy
Effective dose
Male:
Female:
2.58 mSv
2.79 mSv
3.09 mSv
3.23 mSv