Dose Information
– if the automatic dose increase for obese patients (or
patient sections) has to be more moderate than the
preset (choose obese: weak increase), resulting in
more image noise and a lower dose for those
images.
– if the automatic dose decrease for slim patients (or
patient sections) has to be stronger than the preset
(choose slim: strong decrease), resulting in more
image noise and a lower dose for those images.
– if the automatic dose decrease for slim patients (or
patient sections) has to be more moderate than the
preset (choose slim: weak decrease), resulting in
less image noise and a higher dose for those images.
On the Patient tab card you can adjust the image qual-
ity (for more information see chapter How does CARE
Dose 4D work).
Note: Changing this adaptation strength effects all
protocols!

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV
Scan Time per Rotation0.25 seconds
SoftwareSyngo
ConnectivityDICOM

Related product manuals