Table 4-4 Analog Filter Error in DCI measurement
Range
Reading Rate
Additional Analog Filter Error
200μA
Slow
0.002% range
Middle
0.005% range
Fast
0.005% range
20mA2A
Slow
0.040% range
Middle
0.060% range
Fast
0.080% range
200mA
Slow
0.004% range
Middle
0.010% range
Fast
0.010% range
10A
Slow
0.008% range
Middle
0.010% range
Fast
0.010% range