SDM3045X Digital Multimeter 37
Press All, the measured value will be shown on the Vice display and
the corresponding value will be shown on the Main display at the same
time, which is convenient for users to observe temperature and voltage
values ( or resistance value).
Diagram 2- 32 Show Temperature And Voltage Values
6. Press Unitsto choose the unit of temperature. The Multimeter supports
three units: °C , °F, K.
Diagram 2- 33 Unit Selection Interface
7. Set relative value (Optional operation).
Press Rel to open or close Relative math function. When it is opened,
the reading displayed is a value which comes from the result of actual
measurement value subtracts the relative value that has been set. The
default relative value is the measurement value when the function is
opened. (Please refer to Math Functions in Chapter 2 to know about the
details.)
8. Read measurement result.
Temperature measurement is fixed at Fast rate. Therefore, you can‟t
adjust the reading rate when reading the result.