Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
3 Safety
Version 03/2014, Rev. 2.22 of 24/10/2023 sb
Original operating manual, part no. 0700902
Pos: 42 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0060 Sicherheitshinw eise------------------------------------------------------------- @ 26\mod_140531902 1480_68.docx @ 192031 @ 2 @ 1
3.6 Safety instructions
Pos: 43 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0060-0 010 Elektrisch e Sicherheit @ 67\mod_ 1569474663546_6 8.docx @ 716 977 @ 3 @ 1
3.6.1 Electrical safety
As protection against electric shock, the centrifuge is equipped with an
earthed mains power cable and connector. To ensure the effectiveness of
this safety feature, the following must be ensured:
DANGER
Ensure that the wall socket is properly wired and grounded.
Check that the mains voltage agrees with the nominal voltage listed on
the name plate.
Ensure that the mains power cable is intact prior to using the
centrifuge. Damaged or faulty mains power cables must be replaced
immediately.
Do not place vessels containing liquid on the centrifuge lid or within
the safety distance of 30 cm around the centrifuge. Spilled liquids may
get into the centrifuge and damage electrical or mechanical
components.
Only qualified and specialised personnel are authorised to perform
service tasks or repairs of the electrical system for which the housing
needs to be removed.
Inspect the electrical equipment of the unit regularly. Defects such as
loose or burnt cables must be eliminated immediately.
Following the completion of any type of repair or service, the qualified
and specialised personnel must perform final inspection and testing in
compliance with the relevant standards.
Pos: 44 /010 Univers almodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 45 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0060-0 020 Mecha nische Sicherh eit (außer 1-1 6/1-16K/1-7) @ 2 6\mod_14053 19024314_68.d ocx @ 192059 @ 3 @ 1
3.6.2 Mechanical safety
In order to ensure the safe operation of the centrifuge, observe the
following:
WARNING
Do not open the lid when the rotor is in motion!
Do not reach into the rotor chamber when the rotor is in motion!
Do not use the centrifuge if it was installed incorrectly.
Do not use the centrifuge without panels.
Do not use the centrifuge if the rotors and inserts show signs of
corrosion or other defects.
Only use the centrifuge with rotors and accessories that have been
approved by the manufacturer. In case of doubt, contact the
manufacturer (see chapter 7.3 - "Service contact").
Do not hold your fingers between the lid and the housing when closing
the lid. Risk of crushing!
Defective lid relieving devices could cause the centrifuge lid to fall
(contact the service department, if necessary). Risk of crushing!
Do not hit or move the centrifuge during its operation.
Do not lean against or rest on the centrifuge during its operation.