Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
7 Malfunctions and error correction
Version 03/2014, Rev. 2.22 of 24/10/2023 • sb
Original operating manual, part no. 0700902
Pos: 166 /100 Sigm a/110 BA Ze ntrifugen Sigma (Projekte)/3-1 6L/070 Störu ngen und Fehlersu che/070- 0010-0010 Noten triegelung 3-16 L @ 29\mod_ 1405321641164_ 68.docx @ 197 281 @ 3 @ 1
7.1.1 Emergency lid release
In the event of a power failure, it is possible to manually open the centrifuge
lid.
• Switch off the mains power switch and disconnect the power cord from
the socket.
• Remove the stopper (see figure, item 1) from the opening at the side
panel, e.g. with a screwdriver.
Fig. 31: Position of the opening for the emergency lid release
• First, pull the red string to release the mechanism and then unlock the
lid lock by pulling the white string.
• Then, reinsert the plug and quit the error message by switching the
centrifuge off/on. Approximately 10 seconds after quitting the lid can be
closed again.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.