6 Using the centrifuge
Version 03/2014, Rev. 2.22 of 24/10/2023 sb
47 / 84
Original operating manual, part no. 0700902
Pos: 153 /010 Univ ersalmodule/ Seitenwechs el @ 0\mo d_120211624 4312_0.docx @ 105 @ @ 1
Pos: 154 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/060 Betrieb _Spincontrol L+S/Spincontrol L (Z ent6)/060- 0030-0030 Programmbetri eb -ohne Tem p @ 27\mod_140 5319308365_ 68.docx @ 193 011 @ 3 @ 1
6.3.3 Program mode
A program contains all the data that are required for a centrifuge run.
Certain sedimentation results can be repeated under identical conditions.
Programs can be saved, loaded, executed, and edited, when the centrifuge
is at a standstill.
A maximum of 50 programs can be stored under the numbers 1 - 50.
"--" means that the values that are currently set are not a stored program.
The programs can be protected against unauthorised use, modification, or
deletion with the aid of an input lock (see chapter 6.3.2.13 - "Input lock").
Pos: 155 /010 Univ ersalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 156 /100 Sigm a/100 BA Ze ntrifugen Sigma (Standardm odule)/060 Betr ieb_Spincontrol L+S/Spinco ntrol L (Zent6)/060- 0030-0030- 0011 Programm speichern - ohne Temp @ 2 7\mod_14053 19311134_68.d ocx @ 193039 @ 4 @ 1
6.3.3.1 Saving a program
NOTE
This function is only available if the centrifuge is at a standstill.
Enter the parameters that are to be included in the program.
Select the menu item "save Prog" and confirm the selection. The
indication "save" flashes in the activated status.
Select a random storage location from the program selection list. Free
storage locations are indicated by a flashing display. Any storage
locations that are already occupied will be overwritten during the saving
process.
Save the program in the desired location.
The program is now saved.
NOTE
The functions "Auto Open", "Buzzer", "ProgLock", and "Cycles" cannot be
entered as part of a program.
Fig. 29: Saving a program
Pos: 157 /010 Univ ersalmodule/ Seite nwechsel @ 0\mo d_120211624 4312_0.docx @ 105 @ @ 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma 3-16L and is the answer not in the manual?

Sigma 3-16L Specifications

General IconGeneral
BrandSigma
Model3-16L
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals