Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
6 Using the centrifuge
Version 03/2014, Rev. 2.22 of 24/10/2023 sb
Original operating manual, part no. 0700902
Pos: 118 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/060 Betrieb _Spincontrol L+S/060-002 0-0020-0030 Einsetzen von Z ubehör 4plätzig 2-16P, 2-16KL, 3-16L, 3-16KL , 3-18KS @ 63\mod_1554 803016350_68. docx @ 683660 @ 4 @ 1
6.2.2.3 Installation of accessories
Only use vessels that are suitable for the rotor.
In the case of swing-out rotors, all of the compartments must be
equipped with buckets.
Always load opposite compartments of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with vessels of various sizes
Working with vessel of various sizes is possible. In this case, however, it is
very important that the rotationally symmetrical inserts are identical.
correct
incorrect
Fig. 11: Permissible and impermissible loading of a swing-out rotor with vessels of various
sizes (example illustration)
Centrifugation with low capacity
Install the sample vessels in a rotationally symmetrical manner so that
the buckets and their suspensions are loaded evenly.
It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 12: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)