Pos: 118 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/060 Betrieb _Spincontrol L+S/060-002 0-0020-0030 Einsetzen von Z ubehör 4plätzig 2-16P, 2-16KL, 3-16L, 3-16KL , 3-18KS @ 63\mod_1554 803016350_68. docx @ 683660 @ 4 @ 1
→
6.2.2.3 Installation of accessories
• Only use vessels that are suitable for the rotor.
• In the case of swing-out rotors, all of the compartments must be
equipped with buckets.
• Always load opposite compartments of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with vessels of various sizes
Working with vessel of various sizes is possible. In this case, however, it is
very important that the rotationally symmetrical inserts are identical.
Fig. 11: Permissible and impermissible loading of a swing-out rotor with vessels of various
sizes (example illustration)
Centrifugation with low capacity
• Install the sample vessels in a rotationally symmetrical manner so that
the buckets and their suspensions are loaded evenly.
• It is not permissible to load angle rotors on only one axis.