8 Maintenance and service
Version 03/2014, Rev. 2.22 of 24/10/2023 sb
55 / 84
Original operating manual, part no. 0700902
Pos: 182 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/080 Wartu ng und Instandh altung/080-0 010-0030 Rot or, Becher u nd Vielfachträger ( außer 1er B asic) @ 28\mod_ 14053206210 63_68.docx @ 1 94873 @ 3 @ 1
8.1.3 Rotors, buckets and carriers
Rotors, buckets and carriers are produced with the highest precision, in
order to withstand the permanent high stress from high gravitational fields.
Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metals. Barely
detectable cracks on the surface can expand and weaken the material
without any visible signs.
Check the material regularly (at least once a month) for
cracks
visible damage of the surface
pressure marks
signs of corrosion
other changes.
Check the bores of the rotors and multiple carriers.
Replace any damaged components immediately for your own safety.
After every cleaning process, grease the rotor tie-down screw slightly
with a small amount of heavy-duty grease for load-bearing bolts (part
no. 71401) and distribute the grease with a cloth so that it forms a thin
layer.
Pos: 183 /010 Univ ersalmodule/ Seitenwechs el @ 0\mo d_120211624 4312_0.docx @ 105 @ @ 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma 3-16L and is the answer not in the manual?

Sigma 3-16L Specifications

General IconGeneral
BrandSigma
Model3-16L
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals