5 Set-up and connection
Pos: 98 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/050 Aufstellu ng und Anschl uss/050-001 0 Aufstellort (a ußer 8KS) @ 2 7\mod_14053 19174276_68.d ocx @ 192395 @ 2 @ 1
→
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to the ambient air.
• Ensure sufficient ventilation.
• Keep a safety range of at least 30 cm free around the centrifuge as well
as with regard to walls or other devices so that the vents in the machine
remain unobstructed and fully effective.
• Do not subject the centrifuge to thermal stress, e.g. by positioning it
near heat generators.
• Avoid direct sunlight (UV radiation).
• The table must be stable and have a solid, even surface.
• Attention: During transport from cold to warmer places, condensational
water will collect inside the centrifuge. It is important to allow sufficient
time for drying (min. 24 h) before the centrifuge can be used again.
Pos: 99 /010 Univers almodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 100 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/050 Aufst ellung und Anschlu ss/050-00 20 Energiev ersorgung------------------------------------------------------ @ 27\mod_14053191756 30_68.docx @ 192409 @ 2 @ 1
→
5.2 Power supply
Pos: 101 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/050 Aufst ellung und Anschlu ss/050-00 20-0011 Ansc hlussart (Kaltgerä testecker+Sch utzschalter) @ 32\mod_141 7600691402_ 68.docx @ 222 498 @ 3 @ 1
5.2.1 Type of connection