8.1.2 Accessories
For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time!
• Immediately rinse off the rotor, buckets, or accessories under running
water if they have come into contact with any liquids that may cause
corrosion. Use a brush for test tubes in order to clean the bores of angle
rotors. Turn the rotor upside down and allow it to dry completely.
• Clean the accessories outside the centrifuge once a week or preferably
after each use. Adapters should be removed, cleaned and dried.
Do not clean the accessories in a dishwasher!
Cleaning in a dishwasher removes the anodised coating; the result is
cracking in areas that are subject to stress.
• If the rotors or accessories have been contaminated with toxic,
radioactive, or pathogenic substances, clean them immediately with a
suitable decontamination agent (depending on the type of
contamination). Take suitable precautions for your own safety if there is
a risk of toxic, radioactive, or pathogenic contamination.
• Dry the accessories with a soft cloth or in a drying chamber at
approx. 50°C.
Pos: 179 /010 Univ ersalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 180 /100 Sigm a/100 BA Ze ntrifugen Sigma ( Standardmo dule)/080 Wartu ng und Instandh altung/080-0 010-0020-0 010 Kunststoffz ubehör @ 28\mo d_1405320 619602_68.docx @ 194859 @ 4 @ 1
→
8.1.2.1 Plastic accessories
The chemical resistance of plastic decreases with rising temperatures (see
chapter 11.5 - "Resistance data").
• If solvents, acids, or alkaline solutions have been used, clean the plastic
accessories thoroughly.
Plastic accessories must not be greased!