Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
5 Set-up and connection
Version 03/2014, Rev. 2.22 of 24/10/2023 sb
Original operating manual, part no. 0700902
The centrifuges are equipped with a mains power switch with an integrated
thermal circuit breaker.
Switch the unit off by actuating the mains power switch.
If it has tripped, let the circuit breaker cool for approximately 2 minutes.
Switch the unit on.
The centrifuge is now ready for operation.
Pos: 102 /010 Univ ersalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 103 /100 Sigm a/100 BA Ze ntrifugen Sigma (Standardm odule)/050 Aufst ellung und Ansc hluss/050-00 20-0020 Sic herungen bausei ts (außer 3-3 0KS, 8KS, 8KBS) @ 7\mod_13 06312302397 _68.docx @ 4351 3 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp B fuses that are to
be provided by the customer.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.
Pos: 104 /010 Univ ersalmodule/ Abschnittsw echsel @ 0 \mod_1202 124514062_0.d ocx @ 418 @ @ 1
Pos: 105 /010 Univ ersalmodule/ Seitenwechs el @ 0\mo d_120211624 4312_0.docx @ 105 @ @ 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma 3-16L and is the answer not in the manual?

Sigma 3-16L Specifications

General IconGeneral
BrandSigma
Model3-16L
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals