11.4 Table of the service life of rotors and accessories
• The rotors and accessories must be put out of service after 10 years. Any use after this period
may be permissible in individual cases after an inspection performed by the manufacturer.
• If a specification concerning the maximum number of cycles and a specification concerning the
service life (i.e. a date) are provided, the specification that occurs first shall apply.
• After 50,000 cycles, rotors must be scrapped for safety reasons.
Service life
("Exp.Date")
Suitable for centrifuge Remarks
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS, 6-16KS, 6-16KHS
4-5KL, 4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS, 6-16KS, 6-16KHS
6-16S, 6-16HS, 6-16KS, 6-16KHS
6-16S, 6-16HS, 6-16KS, 6-16KHS
Do not grease the
load-bearing bolts
of the rotor
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS, 6-16KS, 6-16KHS
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS, 6-16KS, 6-16KHS
2-7, 2-16P, 2-16KL, 2-16KHL
Do not grease the
load-bearing bolts
of the rotor
2-7, 2-16P, 2-16KL, 2-16KHL,
3-30KS, 3-30KHS
Do not grease the
load-bearing bolts
of the rotor
6-16S, 6-16HS, 6-16KS, 6-16KHS
6-16S, 6-16HS, 6-16KS, 6-16KHS
Special software
required