Levelogger Series User Guide
Page 59
Example 8.1 Depth to Water Level Adjustment
When using a manual depth to water measurement taken from the top of a
well casing as a Field Zero, enter it as a positive value (e.g. 10 m).
In your adjusted data file, your readings will increase in value as the water
level decreases. This is because the depth to static water level from the top
of the well casing is increasing. The adjusted readings will decrease in value
as the static water level rises.
Original Levelogger Data
(barometrically compensated
height of water above sensor)
8.75 m 8.50 m 8.75 m 9.0 m 9.25 m
Adjusted Levelogger Data
(depth to water)
(A) 10.0 m
10.25 m 10.0 m 9.75 m 9.50 m