redPOWER QUBE Cased Fiber Laser User Manual
Laser OFF
Laser STANDBY
Laser ON
The states are described in more detail below.
OFF State
The OFF state is entered under the following conditions:
Following the appropriate input to the Machine Interface
Following a Laser OFF command
Following an Alarm condition
The OFF state is left after receiving a Laser STANDBY command.
STARTING State
When in the OFF state, the STARTING state is entered after receiving a Laser STANDBY
command. The STARTING state automatically sequences to the STANDBY state after a
fixed time of 5s.
The OFF state is entered if an Alarm condition occurs.
STANDBY State
The STANDBY state is entered under the following conditions
Following the STARTING state duration after a Laser STANDBY command is
received.
Following the Ramp Down state duration after a Laser STANDBY command is
received.
The STANDBY state is left after receiving a Laser ON command.
The OFF state is entered if an Alarm condition occurs.
RAMPING UP State
When in the RAMPING DOWN or STANDBY states, the RAMPING UP state is entered
after receiving a Laser ON command.
When in other states the Laser ON command is ignored.
The RAMPING UP State automatically sequences to the ON state after the ramp up
duration has elapsed. The ramp up duration is determined by the active parameter set.
The OFF state is entered if an Alarm condition occurs.
ON State
The ON state is entered automatically from the RAMPING UP State after the ramp up
duration has elapsed. The ramp up duration is determined by the active parameter set.
The ON state is left after receiving a Laser STANDBY command.