EasyManua.ls Logo

Storz AIDA C WD310 - Page 101

Storz AIDA C WD310
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electromagnetic compatibility
Instructions for use • AIDA C WD310 • AAB272_EN_V2.0_12-2021_IFU_CE-MDR 101
10.7 Table 5 – Recommended separation distances between
portable and mobile HF communications equipment
and the product
The product is intended for use in an electromagnetic environment in which HF disturbances
are controlled. The customer or user of the product can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile HF
communications equipment (transmitters) and the product as recommended below, according
to the output energy of the communications equipment.
Rated power of the
transmitter [W]
Separation distance d [m] according to the transmitter frequency
150kHz to 80MHz
d = 1.2√P
80MHz to 800MHz
d = 1.2√P
800MHz to 2.5GHz
d = 2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters whose maximum rated power is not listed in the table above, the recom-
mended separation distance d in meters (m) can be estimated using the equation from the re-
spective column, whereby P is the maximum rated power of the transmitter in watts (W) ac-
cording to the transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher frequency range ap-
plies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic
waves is affected by absorptions and reflections of buildings, objects, and people.