30 3006-109-101 REV A www.stryker.com
Return To Table of Contents
LED INDICATORS: FOOTBOARD (CONTINUED)
D
A
B
C
E F G
H
Footboard Operation Guide
LED Name: Function LED Color
A
Bed Motion Lock LED: LED is illuminated when Bed Motion Lock is activated
or when the Patient Control (Fowler, Gatch, Bed Up/Down) Lock buttons are
activated.
AMBER
B
Fowler 30
0
+ LED: LED is illuminated when the Fowler 30+ is locked. The
LED will blink if the iBed
®
Awareness system is “On”, the Fowler 30+ is being
monitored and the Fowler goes below 30 degrees or the Fowler 30+ is turned
Off”.
AMBER
C
Low Height LED: LED is illuminated when bed is in low height. The LED will
blink if the iBed
®
Awareness system is “On”, the low height is being monitored,
and the bed is not in low height.
AMBER
D
Brake LED: LED is illuminated when the brake is set, and will blink if the brake
is not set.
AMBER
E
Bed Exit LED (Optional): LED is illuminated when the Bed Exit is armed. The
LED will blink if the Bed Exit is turned Off while the iBed
®
Awareness system is
turned On or if Bed Exit alarms while monitored by iBed
®
Awareness system.
AMBER
F
Bed Zero LED (Optional iBed
®
Awareness): LED is illuminated if Bed Zero is
successful.
AMBER
G
Siderail LED (Optional iBed
®
Awareness): LED is illuminated if iBed
®
Awareness system is “On”. The LED will blink when siderail state has changed.
AMBER
H
Power LED: LED is illuminated when bed has power. GREEN