40 3006 -109-101 REV A www.stryker.com
Return To Table of Contents
IBED
®
AWARENESS FUNCTIONALITY
• The iBed
®
Awareness provides functionality that will monitor status conditions on the product and produce an alert
if the state had changed.
• When the system is turned “On”, it monitors each of the siderail positions and brake automatically. If the bed is in
Low Height and/or the Bed Exit is armed and/or the Fowler 30
0
+ lock is “On”, the system will also monitor these
features when iBed
®
Awareness is turned “On”. Note: If the Fowler 30
0
+ lock is “On” before iBed
®
Awareness is
“On”, the system will also monitor the Fowler 30
0
+ lock.
• In the event of a power loss, the iBed
®
Awareness system will store the last known condition and when power is
restored it will operate in this condition.
• iBed
®
Awareness will not be able to be turned “On” if any system error conditions exist that impede the function of
the iBed
®
Awareness system. The system errors that affect this feature include the four siderail sensors, the scale
system, the Fowler 30
0
+ lock and the bed exit system. For details on error codes, refer to the Maintenance Manual.
IBED
®
AWARENESS LIGHT BAR AND SIDE LIGHTS
A light bar, located centrally on the front of the footboard, will illuminate and indicate the state of the iBed
Awareness system. Side lights located on the sides of the footboard will behave identical to the center light.
Features
• When the iBed
®
Awareness system is “On” the light bar turns green.
• If an alert state on the iBed
®
Awareness system is triggered, the light bar will change to the alert state and
flash AMBER.
• During an alert state, an AMBER dashboard LED associated with the alert will blink on the footboard and the
display screen will show the details of the alert state.
IBED
®
AWARENESS ON/OFF BUTTON
The On/Off control button is used to turn the iBed
®
Awareness system “On” and “Off”.
Features
When the button is pressed the iBed
®
Awareness system will save information based on the current state of the
product and based on the system rules will commence monitoring.
Turning on the iBed
®
Awareness system
1. Press the iBed On/Off button.
2. The following message will be displayed on the screen: “Awareness On”.
Turning off the iBed
®
Awareness system
1. Press and hold the iBed On/Off button.
2. The following message will be displayed on the screen: “Awareness Off”.
Footboard Operation Guide