144 6390-009-002 REV D www.stryker.com
EMC information
Guidance and manufacturer’s declaration - electromagnetic immunity
Power-LOAD is suitable for use in the electromagnetic environment specified below. The customer or the user of
Power-LOAD should assure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level Compliance level
Electromagnetic
environment-guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Surge
IEC 61000-4-5
+1 kV line(s) to line(s)
+2 kV line(s) to ground
+1 kV line(s) to line(s)
+2 kV line(s) to ground
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electrical transient
conduction along supply
ISO 7637-2
per ISO 7637-2 per ISO 7637-3
N/A
Portable and mobile RF
communications equipment
should be used no closer
to any part of Power-LOAD,
including cables, than the
recommended separation
distance calculated from
the equation appropriate
for the frequency of the
transmitter.
Recommended separation
distance
D=(1.2) (√P)
D=(.35) (√P)
80 MHz to 800 MHz
D=(0.70) (√P)
800 MHz to 2.7 GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
Conducted RF
IEC 61000-4-6
3 Vrms
6 Vrms in ISM and amateur
radio bands
150 kHz to 80 MHz
3 V
6 Vrms in ISM and amateur
radio bands
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m