www.stryker.com 6390-009-002 REV D 73
Service
Primary coil replacement, foot end (Continued)
Figure 42: Unplug coil wires from the inductive primary board
Primary coil replacement, head end
Tools required
• 5/32'' hex wrench
• (2) 1/8'' hex wrench
• 3/8'' hex driver
• T25 Torx driver
• Diagonal pliers
• Torque wrench (ft-lb) > 60 ft-lb
Procedure
1. Remove the trolley. See Trolley removal on page 37.
2. Remove the transfer. See Transfer removal on page 37.
3. Using a 3/8'' hex driver, remove the four anchor mounting bolts that secure the anchor to the floor place.
4. Unplug the main power cable from the patient compartment to the anchor assembly at the head end.
5. Using a 5/32'' hex wrench, remove the four screws (A) (Figure 43 on page 74).
6. Using two 1/8'' hex wrenches, remove the four flat head cap screws (B) that secure the anchor pawl assembly to
the anchor (Figure 43 on page 74).
7. Pull the anchor pawl assembly toward the head end of the anchor to remove the anchor pawl assembly.