www.stryker.com 6390-009-002 REV D 55
Service
Inductive coil replacement (Continued)
5. Remove the return spring (D) (Figure 22 on page 55). Save for reinstallation.
D
Figure 22: Return spring location
6. Using a T25 Torx driver, remove the screws (E) that secure the charge bracket (Figure 21 on page 54).
7. Using a 5/32" hex wrench, remove the screw (F) that secures the charge bracket (Figure 21 on page 54).
8. Using a T25 Torx driver, remove the two screws that secure the back left wing cover.
9. Unplug the cable (G) from the trolley control board, and remove the inductive coil (H) (Figure 21 on page 54).
Discard the inductive coil.
Notes
Lift up on the trolley control board to pull the cable through.
Pay attention to the cable routing for reinstallation.
Do not dispose of as unsorted municipal waste. Refer to your local distributor for return or collection systems
available in your country.
10. Reverse steps to reinstall.
11. Verify proper operation of the product before returning it to service.
Trolley position sensor (TPS) replacement
Tools required
T25 Torx driver
1/8" hex wrench
5/32" hex wrench
1/4" ratchet
Procedure
1. Remove the trolley. See Trolley removal on page 37.
2. Remove the trolley covers. See Cover removal and replacement on page 40.