38 6390-009-002 REV D www.stryker.com
Service
Trolley removal (Continued)
4. Slide the patient right transfer lock trigger (C) to extend (Figure 3 on page 38).
Figure 3: Trolley mid transfer position
5. Pull the transfer toward the foot end until the transfer stops in the first lock position.
6. Using a T25 Torx driver, remove the two screws that secure the transfer trim (E) (Figure 4 on page 39).
Note: The trolley will drop down slightly as you remove it from the transfer. Do not tip the trolley upside down,
forward, or onto its side.