www.stryker.com 6390-009-002 REV D 57
Service
Trolley position sensor (TPS) replacement (Continued)
6. Using a 1/8" hex wrench, loosen (do not remove) the flat head cap screw (E) that secures the front link (F) to the
trolley mechanism arm and swing it out of the way (Figure 24 on page 57).
Note: Rotate the link to remove the pivot pillar. During removal, pay attention to the orientation of the pivot pillar.
B
C
D
E
F
C
E
F
Figure 24: TPS assembly