30.Assemble the bumper block. Align the supplied inner dead stop block (639000010246) (with threaded holes) (Y) with the
supplied dead stop bumper (639000010243) (X) (
Figure 16).
W
X
Y
Z
W
X
Z
FFiigguurree 1166 AAsssseemmbbllee bbuummppeerr bblloocckk
31.Slide the assembled components (X and Y) into the end of the transfer assembly.
NNoottee -- Position the dead stop bumper, so that the hump (AA) is on the bottom (
Figure 17).
AA
FFiigguurree 1177 DDeeaadd ssttoopp bbuummppeerr hhuummpp
32.Align the dead stop block (639000010244) (with thru holes) (W) with the assembled components (X and Y) and slide it
into the end of the transfer assembly.
33.Repeat steps 26 - 28 to assemble and install the second dead stop bumper.
34.Using a 3/16'' hex wrench, install the four supplied socket head cap screws (0004-662-000) (two for each side) (Z) into
the transfer assembly (
Figure 16).
35.Visually inspect the head end transfer bumpers to make sure that they are installed flush to the outer edge of the
transfer assembly with no signs of misalignment or improper installation.
36.Using a T25 Torx driver, install the two supplied button head cap screws (0004-665-000) (AB) to attach the supplied
patient left transfer head end trim (639000010210) and supplied patient right transfer head end trim (639000010211)
(AC) to the transfer assembly (
Figure 18).
EN 22 6390-709-001 Rev AB.1