technical reference
329
Synrad Firestar f-Series operator’s manual
f100 general specications
Parameter Firestar f100
Cooling Specications
Maximum Heat Load ............................................2000 Watts (6.83 kBtu/hr)
Flow Rate (minimum) ..........................................2.0 GPM at < 60 PSI (7.6 lpm at < 414 kPa)
Pressure Drop ..........................................................10 PSI at 2 GPM (69 kPa at 7.6 lpm)
Coolant Temperature
8
..........................................18 °C to 22 °C
Environmental Specications
Operating Ambient Temperature Range
9
.....15 °C to 40 °C
Humidity ...................................................................0–95%, non-condensing
Physical Specications
Length .......................................................................27.4 in (69.6 cm)
Width .........................................................................7.75 in (19.7 cm)
Height ........................................................................5.84 in (14.8 cm)
Weight .......................................................................37 lbs (16.8 kg)
* Specications subject to change without notice.
8 Firestar f100 lasers can be operated at coolant temperatures up to 24 °C (75 °F) in order to reduce problems associated with
condensation; however, this may result in decreased laser performance and/or reduced laser lifetime.
9 Published specications guaranteed at a cooling temperature of 22 °C.